Agilent Technologies B2902 Manual de Serviço Página 183

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 326
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 182
Agilent B2900 SCPI Command Reference, Edition 2 4- 61
Subsystem Commands
:PROGram:CATalog?
PROGram Subsystem
For the numeric suffix [h], see “Numeric Suffix” on page 1-8.
:PROGram:CATalog?
Returns the names of all programs defined in the program memory.
Even if a name is selected by the :PROGram[:SELected]:NAME command, this
command does not return the name if the program is empty.
Syntax :PROGram:CATalog?
Query response program_names <newline>
program_names returns the names of all programs defined in the program memory.
Response data type is AARD.
Example :PROG:CAT?
:PROGram:PON:COPY
Specifies the power-on program.
Syntax :PROGram:PON:COPY name
Parameter name Name of the program used for the power-on program.
Parameter data type is SPD.
Example :PROG:PON:COPY “program1”
:PROGram:PON:DELete
Clears the power-on program.
Syntax :PROGram:PON:DELete
Example :PROG:PON:DEL
Vista de página 182
1 2 ... 178 179 180 181 182 183 184 185 186 187 188 ... 325 326

Comentários a estes Manuais

Sem comentários