418 Chapter7
Language Reference
AMPCORDATA Amplitude Correction Data
The values of the correction points are applied across the active
measurement range. Between points, the correction values are
interpolated. When measuring at frequencies outside the first and last
correction points, these values are used as the correction value. If you
do not want any amplitude correction outside of the first and last
correction points, set the amplitude correction to 0 at the frequencies
that are outside of the first and last correction values.
If any of the trace data is above or below the graticule, the ampcor
function may not properly correct it. The spectrum analyzer amplitude
accuracy is not specified above or below the graticule. Anytime ampcor
applies a correction such that data outside the graticule is moved to
within the graticule, an error (error number 921 or 922) is generated. In
order to avoid these errors, make sure that the trace data which is
being corrected is within the graticule before the correction is applied.
Query Response
Figure 7-52 AMPCORDATA Query Response
Example
10 OUTPUT 718;"AMPCORDATA 1GHZ, -1.56DB, 10MHZ, 10DB;"