90 Chapter2
Making Measurements
Example 7: Stimulus-Response Measurements
Figure 2-35 Guided calibration routines prompt the user.
Figure 2-36 The thru trace is displayed in trace B.
Normalize
Normalization eliminates the frequency response error in the test
setup. When normalization is on, trace math is performed on the active
trace:
A B + NRP A
1. where:
A is the active trace.
B is the stored thru calibration trace.
NRP is the normalized reference position.