Agilent Technologies 54503A Guia do Utilizador Página 52

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 716
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 51
52 Chapter2
Making Measurements
Example 1: Resolving Closely Spaced Signals (with Resolution Bandwidth)
To resolve two signals with a frequency separation of 2 kHz, a 1 kHz
resolution bandwidth again must be used (see Figure 2-2). Since the
spectrum analyzer uses bandwidths in a 1, 3, 10 sequence, the next
larger filter, 3 kHz, would exceed the 2 kHz separation and thus would
not resolve the signals.
Keep in mind that noise sidebands (phase noise) can also affect
resolution.
Figure 2-1 1 kHz Signal Separation
Figure 2-2 2 kHz Signal Separation
Vista de página 51