1-67
Making Measurements
Measuring Amplifiers
With the previous points in mind, the amount of attenuation can be calculated from the
following equations:
Attenuator value = +20 dBm 13 dB (10 dBm). Attenuator Value = +17 dB
3. Choose the S21 measurement parameter by pressing .
4. Connect the attenuator between the B OUT and IN ports.
5. Perform a calibration.
6. Connect the DUT as shown in Figure 1-53 and view the measurement.
Figure 1-53 Option 012 High Power Configuration
Meas
Trans: FWD S21 (B/R)