Making Measurements
Measuring Electrical Length and Phase Distortion
Group delay is calculated by dividing the phase difference between points by the frequency
spacing. Thus, if n equals the number of points, the number of phase difference values (or
frequency segments) will be n1. The first data point is repeated so that the total number
of points remains n.