Agilent Technologies 24A Guia do Utilizador Página 320

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 453
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 319
6-32
Calibrating for Increased Measurement Accuracy
Full Two-Port Error Correction (ES Analyzers Only)
17.To compute the error coefficients, press:
The analyzer displays the corrected measurement trace. The analyzer also shows the
notation Cor at the left of the screen, indicating that error correction is on.
NOTE You can save or store the measurement correction to use for later
measurements. Refer to Chapter 4 , “Printing, Plotting, and Saving
Measurement Results” for procedures.
18.This completes the full two-port correction procedure. You can connect and measure
your device under test.
DONE 2-PORT CAL
Vista de página 319
1 2 ... 315 316 317 318 319 320 321 322 323 324 325 ... 452 453

Comentários a estes Manuais

Sem comentários