2-47
Making Mixer Measurements (Option 089 Only)
Isolation Example Measurements
4. To select a ratio B/R measurement, press:
NOTE Isolation is dependent on LO power level and frequency. To ensure good test
results, you should choose these parameters as close to actual operating
conditions as possible.
5. Make the connections as shown in Figure 2-37.
Figure 2-37 Connections for a Response Calibration
6. Perform a response calibration by pressing
.
7. Make the connections as shown in Figure 2-38.
Figure 2-38 Connections for a Mixer RF Feedthrough Measurement
Meas
INPUT PORTS
B/R
Cal
CALIBRATE MENU
RESPONSE
THRU