7-65
Operating Concepts
Modifying Calibration Kits
NOTE The published specifications for this network analyzer system include
accuracy enhancement with compatible calibration kits. Measurement
calibrations made with user-defined or modified calibration kits are not
subject to the analyzer specifications, although a procedure similar to the
system verification procedure may be used.
Saving Modified Calibration Kits to a Disk
The calibration kit, along with any calibration data and other instrument state
information, can be saved to an ISTATE file on a floppy disk. To save a modified calibration
kit with an instrument state, press:
Cal
CAL KIT
SELECT CAL KIT
USER KIT
Save/Recall
SAVE STATE
Vista de página 409