218 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
24 Calibrating the Infiniium Oscilloscope and Probe
Figure 83 Time Scale Calibration Dialog box
d Click on the Std + Dflt button to continue the calibration, using the
Factory default calibration factors.
e When the calibration procedure is complete, you will be prompted
with a Calibration Complete message window. Click the OK button to
close this window.
f Confirm that the Vertical and Trigger Calibration Status for all
Channels passed.
g Click the Close button to close the calibration window.
h The internal calibration is completed.
i Read NOTE below.
NOTE
These steps do not need to be performed every time a test is run. However, if the ambient
temperature changes more than 5 degrees Celsius from the calibration temperature, this
calibration should be performed again. The delta between the calibration temperature and
the present operating temperature is shown in the Utilities>Calibration menu.