Agilent Technologies DisplayPort Manual de Serviço Página 105

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 246
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 104
Source Main Link Frequency Compliance Differential Tests 12
Agilent U7232A DisplayPort Electrical Performance Compliance Test Application 105
.
You can use any oscilloscope channel and connect it to any lane test point.
You select the channels used for testing lanes in the Set Up tab of the
DisplayPort Electrical Performance Compliance Test Application.
The data lanes and channels shown in the previous figures are just
examples. You can choose any desired data lane and channel that you
want.
Infiniium Oscilloscope
Data Channel
SMA to SMP cables
DUT
Lane on the W2641A test fixture
Figure 35 Differential Measurement Setup Using Two Single Ended Connections - Main
Link Frequency Compliance Tests (A Minus B Configuration)