Agilent Technologies DisplayPort Manual de Serviço Página 217

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 246
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 216
Calibrating the Infiniium Oscilloscope and Probe 24
Agilent U7232A DisplayPort Electrical Performance Compliance Test Application 217
Figure 82 Oscilloscope Calibration Window
5 Follow the on- screen instructions:
a You will be prompted to disconnect everything from all the inputs:
click the OK button.
b Then you will be prompted to connect the calibration cable with
SMA adapters between the Aux Out and a specified input. Install the
SMA adapter by pressing it on input BNC, and hand tightening the
outer ring turning right. Click the OK button after connecting the
cable as prompted.
c During the calibration of channel 1, you will be prompted a
Calibration Options screen, as shown in
Figure 83 below.
Vista de página 216
1 2 ... 212 213 214 215 216 217 218 219 220 221 222 ... 245 246

Comentários a estes Manuais

Sem comentários