Agilent Technologies DisplayPort Manual de Serviço Página 174

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 246
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 173
174 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
20 Sink Non-ISI Jitter Tests
Probing for Sink Non-ISI Jitter Tests
When performing the sink non- ISI jitter test, the DisplayPort Electrical
Performance Compliance Test Application will prompt you to make the
proper connections. Your DisplayPort test environment setup on the Set
Up tab must match the physical connection.
Figure 59 and Figure 60 show a physical connection for making
differential and single- ended connections.
.
Infiniium Oscilloscope
1168A or 1169A
Probe Amplifier
N5380A SMA probe head.
Connect the plus side of the
probe to the plus side of the
SMA probe head and the
negative side of the probe to
the negative side of the SMA
probe head.
Lane on the W2641A test fixture
SMA to SMP cables
DUT
Figure 59 Probing for Differential Tests - Sink Non-ISI Jitter Tests (Single Connection with
W2641A DisplayPort Test Fixture)
Vista de página 173
1 2 ... 169 170 171 172 173 174 175 176 177 178 179 ... 245 246

Comentários a estes Manuais

Sem comentários