Agilent Technologies C-Size VXIbus Systems Manual do Utilizador Página 98

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 178
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 97
2-Frame System - Agilent RADI-EPC7 Embedded
Computer
MAINFRAME 1: Set the VXI-MXI module to address 1. Set other
VXI modules in the mainframe to addresses below 128. Note - The
EPC7 is installed in slots 1 and 2 and is configured as a non-slot 0
device.
MAINFRAME 2: Set the VXI-MXI module address to address
128.
Note - In two and three mainframe systems, the EPC7 Start Up
Resource Manager (SURM) does not allow statically configured
modules to be located at logical address 255.
VXI-MXI Module
LADDR = 128
Other Modules
LADDR’s 129 - 191
EPC7 Computer
LADDR = 0
VXI-MXI Module
LADDR = 1
Other Modules
LADDR’s 2 - 127
Agilent E1482
Agilent E1482
Mainframe 1
(root)
Mainframe 2
(extender)
Agilent RADI-EPC7
Embedded Controller
Logical Address = 0
MXIbus Cable
INTX Cable
Figure 3-16. Embedded EPC7 System, 2-Frame
3-16 Set Up the System for Multiple Mainframes C-Size Configuration Guide
Vista de página 97
1 2 ... 93 94 95 96 97 98 99 100 101 102 103 ... 177 178

Comentários a estes Manuais

Sem comentários