Agilent Technologies C-Size VXIbus Systems Manual do Utilizador Página 100

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 178
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 99
2-Frame System - Agilent V743 Embedded
Controller
MAINFRAME 1: Set the VXI-MXI module to address 1. Set other
VXI modules in the mainframe to addresses below 128. Note: The
resource manager must be at address 0.
MAINFRAME 2: Set the VXI-MXI module address to address
128.
VXI-MXI Module
LADDR = 128
Other Modules
LADDR’s 129 - 255
Embedded Controller
LADDR = 0
VXI-MXI Module
LADDR = 1
Other Modules
LADDR’s 2 - 127
Agilent E1482
Agilent E1482
Mainframe 1
Mainframe 2
Embedded Computer
Logical Address = 0
MXIbus Cable
INTX Cable
Figure 3-18. Embedded V743 2-Frame
3-18 Set Up the System for Multiple Mainframes C-Size Configuration Guide