Agilent Technologies B2201A Manual de Serviço Página 58

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 124
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 57
Module 11
Advanced Definitions and Operations
11-16
Final_xxxx.exe
Start
End
Chuck down
Arguments: -a GPIB_address
-l log_file_name
Example:
–a GPIB0::5::INSTR –l C:\temp\prb.log
Response: none
C:\Program Files\Agilent\B1500\EasyEXP
C:\Program Files\Agilent\B1500\EasyEXP
C:\Program Files\Agilent\B1500\EasyEXP
5
Stop condition
To specify the final procedure on the Repeat Measurement Setup dialog box, enter the full path name
of Final_xxxx.exe into the Final Procedure field.
The procedure is called after one of the stop condition is detected. This procedure sets wafer chuck
to the DOWN position.
Note: To ignore the Arguments, set the GPIB address and log file name in the prober_info.ini.
Vista de página 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 123 124

Comentários a estes Manuais

Sem comentários