Module 13
SPGU Control and Applications
13-36
Class Exercise
Monitor ALWG output by the sampling measurement.
1. Click I/V-t Sampling icon and open setup editor.
2. Define the measurement setup with the ALWG
output.
3. Save the test setup.
4. Connect a device (511 kohm resistor) as shown in the
next page.
5. Perform the test.
Channel Setup:
SMU1 -> V1, I1, Mode=I
SMU2 -> V2, I2, Mode=COMMON
SPGU1 -> SPGUV1, Mode=ALWG
Measurement Setup:
Interval=2 ms
No of Samples=201
SMU1 -> Source=0 A, Compliance=2 V
Range:
SMU1 -> Mode=FIXED, Range=2 V
ADC/Integ:
SMU1 -> ADC=HS ADC
Advanced:
SMU1 -> SMU Filter=OFF
SPGU ALWG Setup:
SPGU1 -> LoadZ=511 kohm
ALWG Source -> Define the waveform as you like.
Display Setup:
X-axis -> Time
Y1-axis -> V1
For the example setup, open the “ALWG monitor” test setup.