Agilent Technologies ESG Guia do Utilizador Página 207

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 206
Chapter 6 191
BERT Testing
Bit Error Rate Tester–Option UN7
In this example, the triggering sequence is where the trigger delay is active with a cycle
count.
The reference data is generated by stored data bits. If the BER measurement accepts data bits
immediately after receiving a trigger, set the trigger delay to On and the trigger delay count to
a value corresponding to the data format. For PN9 set the delay to 9. If the cycle count is set to
more than 1, it is not necessary to store data bits and no unnecessary delay occurs. Refer to
Figure 6-27 and “Repeat Measurements” on page 193.
Figure 6-27
Vista de página 206
1 2 ... 202 203 204 205 206 207 208 209 210 211 212 ... 501 502

Comentários a estes Manuais

Sem comentários