Agilent Technologies E5500 Phase Noise Measurement System
13-5
AM Noise Measurement Examples
AM Noise using an Agilent/HP 70420A Option 001
6. Choose the Sources tab from the Define Measurement window.
a. Enter the carrier (center) frequency of your UUT. Enter the same
frequency for the detector input frequency.
7. Choose the Cal tab from the Define Measurement window.
a. Select Use automatic internal self-calibration as the calibration
method. For more information about various calibration techniques,
refer to Chapter 12, “AM Noise Measurement Fundamentals”.