7-56
Agilent Technologies E5500 Phase Noise Measurement System
Absolute Measurement Examples
RF Synthesizer using DCFM
Figure 7-13 Connect Diagram for the RF Synthesizer (DC FM) Measurement
4. Refer to the following system connect diagram examples for more
information about system interconnections:
“E5501A Standard Connect Diagram Example” on page 7-58
“E5501B Standard Connect Diagram Example” on page 7-59
Internal AM Detector 0 to +20 dBm
Downconverters:
Agilent/HP 70422A 0 to +30 dBm
Agilent/HP 70427A +5 to +15 dBm
CAUTION:
To prevent damage to the Agilent/HP 70420A test set’s hardware components,
the input signal must not be applied to the test set’s signal input connector until
the input attenuator (Option 001) has been correctly set by the phase noise
software, which will occur at the connection diagram.
Characteristics:
Input Impedance 50 ohm Nominal
AM Noise dc coupled to 50 ohm load
Table 7-11 Agilent/HP 70420A Test Set Signal Input Limits and Characteristics