Agilent Technologies E5503A Guia do Utilizador Página 334

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 465
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 333
Agilent Technologies E5500 Phase Noise Measurement System
12-13
AM Noise Measurement Fundamentals
Method 2: Double-Sided Spur
NOTE The carrier-to-sideband ratio for AM is:
Figure 12-12 Measuring the Carrier-to-Sideband Ratio
4. Reconnect the AM detector and enter the carrier-to-sideband ratio and
modulation frequency.
Figure 12-13 Measuring the Calibration Constant
5. Measure the AM detector calibration constant.
6. Turn off AM.
7. Measure noise data and interpret the results.
NOTE The quadrature meter should be at zero volts due to the blocking capacitor at
the AM detectors output.
C
sb
-----
C
sb
-----
20
percentAM
100
------------------------------


log 6dB==
Vista de página 333
1 2 ... 329 330 331 332 333 334 335 336 337 338 339 ... 464 465

Comentários a estes Manuais

Sem comentários