Agilent Technologies E5500 Phase Noise Measurement System
7-49
Absolute Measurement Examples
RF Synthesizer using DCFM
Defining the
Measurement
1. From the File menu, choose Open.
2. If necessary, choose the drive or directory where the file you want is
stored.
3. In the File Name box, choose “RFSynth_DCFM.pnm”.
4. Click the Open button.
The appropriate measurement definition parameters for this example
have been pre-stored in this file. Table7-12onpage7-70lists the
parameter data that has been entered for the RF Synthesizer using
DCFM measurement example.)
NOTE Note that the source parameters entered for step 2 in Table 7-12 may not be
appropriate for the reference source you are using. To change these values,
refer to Table7-10onpage7-50, then continue with step “a”. Otherwise, go
to “Beginning the Measurement” on page 7-55: