13
Reference Specifications
The accuracy needs for testing GSM radios require the unit
to be operated with the high stability reference (Option 001)
or an external high stability reference.
Accuracy (after warm up): ±[(Time since calibration x
aging rate) + temperature effects + accuracy of calibration]
External reference input frequency: 13, 10, 5, 2 or 1 MHz,
±30 ppm
Level: 0 to +10 dBm
Supplemental characteristics
Nominal impedance: 50 Ω
10 MHz out (rear panel BNC) level: > +8.0 dBm nominal
Impedance: 50 Ω nominal
13 MHz out (rear panel BNC) level: > +8.0 dBm nominal
Impedance: 50 Ω nominal
Fixed Reference Mode
Aging: < 2 ppm/year
Temperature stability: ±1 ppm (0° to 55°C)
Warm-up time: < 30 minutes, ±2 ppm of final frequency
Tunable Reference Mode
This allows offsetting the internal reference by a selected
amount relative to the high stability reference (Option 001)
or an external reference.
Required external reference accuracy: ±0.5 ppm
Tune range: ±30 ppm
Reference accuracy: ±1 ppm + accuracy of external
reference or high stability (Option 001)
Temperature stability: ≤4 ppm, for selected offsets of up
to ±30 ppm
General Specifications
Size: 8922 + 83220: 310H x 426W x 574D (mm)
(12.25 x 16.75 x 23 inches)
Weight: 8922 + 83220: 48.8 kg, 107 lbs
Operating temperature: 0° to +55°C
Storage temperature: -40° to +70°C
Power: 100, 120, 220, 240 Vac, 48 to 440 Hz ±12% of line
voltage, approximately 640 VA
Safety (8922): Certified to IEC 348:1978 and
CSA-22.2, No. 231 Series M89
Safety (83220): Certified to IEC 348:1978 and CSA
Bulletin 556B
EMI: Meets the requirements of the European,
EMC directive 89/336/EEC
Video output: The video out connector on the rear
panel outputs a 15 kHz PAL CVBS under-scanning
compatible signal
Reference and General Specifications
Comentários a estes Manuais