0.3 GMSK Data Recovery Specifications
(8922M only)
6
After one timeslot, 577 µs, from an isolated receiver
hop trigger in the GSM frequency bands.
Input frequency setting error: ±100 Hz
Required input phase accuracy: ≤5° rms, ≤20° peak
Demodulation duty cycle: 1 timeslot per frame
Outputs: Data, clock and data valid
Data output clock: Clocked at 1 MHz rate
Delay, data: ≤1 frame (4.62 ms)
Output level: TTL
FM Demodulation Output Specifications
(8922M only)
6
Sensitivity: 20 µV/Hz ±5% (into an open circuit)
Input frequency setting range: ±50 kHz, with
≤100 kHz peak deviation
Input level setting range: ±3 dB
Supplemental characteristics
3 dB bandwidth: DC to 270 kHz
Output impedance: 600 Ω
DC offset: ≤5 mV
Pulse Demodulation Output Specifications
(8922M only)
6
Input frequency setting range: ±50 kHz
Input level setting range: ±3 dB
Rise time (10 to 90%): ≤2.5 µs
Output RF Spectrum Measurement
Specifications
(Requires Option 006)
After one timeslot, 577 µs, from an isolated receiver hop
trigger in the GSM frequency bands.
Input levels for optimum dynamic range
(RF In/Out connector):
+7, +17, +27, +37 dBm
(from 880 to 960 MHz)
-3, +2, +7, +12, +17, +22, +27, +32 dBm
(from 1710 to 1990 MHz)
Input frequency setting range: ±10 kHz
Input level setting range: ±3 dB
Supplemental characteristics
Log linearity: ±0.4 dB
Amplitude flatness: ±1.0 dB
Amplitude resolution: 0.4 dB
Dynamic range (dB): This describes the spectrum analyzer
resolution bandwidth filter used when measuring output
RF spectrum. The dynamic range of the measurement will
be a combination of this filter response and the modulation
spectrum of the incoming signal.
Offset (kHz)
100 200 300 400 600 800 to 1800
Range (dB) 24 42 53 60 63 64
When using output RF spectrum due to the ramping
measurement, the dynamic range is decreased by 12 dB
(due to peak hold).
0.3 GMSK Data Recovery, FM Demodulation Output, Pulse
Demodulation Output, and Output RF Spectrum Measurement
Specifications
6. 8922P, 8922X systems