Signal Generator
Using the Signal Generator
5 Select the channel standard
from the list using the up and
down arrow buttons.
6 [Select]
7 On [Units] select
Chan. Each time you press this menu key, the
analyzer switches between Channel (Chan)
and Frequency (Freq). The Units settings are
coupled across multiple tools.
8 [Fr/Time Ref] Complex signals such as cdma and 1xEV-DO
require accurate time and frequency
references before they can be reliably
generated or demodulated.
9 [GPS] or [External Even Sec] Only
GPS and External Even Second
provide the timing information necessary to
generate an accurately modulated signal.
Select GPS if you are receiving GPS location
information (
System, [Fr/Time Ref
Location], [Show GPS Location] set to
On).
Otherwise, select External Even Second, and
connect your external even second reference
signal to the
Even Second Sync In port on
your test set. It is worth noting that the signal
may disappear momentarily when you switch
between different time references because
the signal needs to be realigned with the
system time.
If you are not needing a time aligned signal
(for example, you are simply generating a
carrier wave), you do not need to use either
GPS or External Even Second Clock as your
time reference.
If you are using a frequency only reference,
the signal will be modulated with the same
accuracy as when using a time reference. The
difference is that it will not be time aligned to
the system time during playback
Step Notes