Agilent Technologies E7495A RF Manual de Serviço Página 202

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 686
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 201
202 Chapter 7
CDMA Analyzer
Optional CDMA Tx Analyzer Measurement Settings
CDMA Analyzer
Setting the Measurement Time Offset Limits
4 Set the upper limit.
a. [High Limit]
b. Enter the number on
the numeric keypad.
c. [GHz], [MHz],
[kHz], or [Hz]
5 Set the lower limit.
a. [Low Limit]
b. Enter the number on
the numeric keypad.
c. [GHz], [MHz],
[kHz], or [Hz]
Step Notes
1 On [Time Offset Limits]
select
On or Off.
Enables you to toggle the measurement limits
function On (active) and Off (inactive).
If set to
On, the displayed time offset metrics
will change from blue to red (fail) or green
(pass) and indicate either a “P” (pass) or “F”
(fail). If the measured values exceed the upper
or lower limit, the words ‘ALL PASS’, in the
bar at the bottom of the screen, will change to
‘FAILURE’.
If set to
Off, the time offset value will be
displayed in blue. The bar at the bottom of the
screen will indicate measurement in progress.
Step Notes
Vista de página 201