Agilent Technologies E7495A RF Manual de Serviço Página 319

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 686
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 318
Chapter 11 319
E1 Analyzer
Measurement Results
E1 Analyzer
Recv Level - The voltage level of the signal measured at the Rx jack.
Frequency - The frequency of the signal measured at the Rx jack. The E1 Pulses
indicator must be active for this measurement to be displayed.
Elapsed Time - Indicates the amount of elapsed time since the last measurement
reset.
BPV Errors - A tally of the number of BPV errors since the last measurement
reset.
BPV Error Rate - The percent ratio of BPV errors to total bits transmitted since
the last measurement reset.
BPV Errd Secs - The number of one second intervals since the last
measurement reset that contained BPV errors.
Frame Errors - A tally of the number of Frame errors since the last
measurement reset.
Frame Error Rate - The percent ratio of frame errors to total bits transmitted
since the last measurement reset.
Frame Errd Secs - A tally of the number of one-second intervals since the last
measurement reset that contained frame errors.
Signal Loss - A tally of the number of one-second intervals since the last
measurement reset that contained signal errors.
Frame Sync Loss - A tally of the number of one-second intervals since the last
measurement reset that contained frame sync loss.
Excess Zeros - A tally of the number of one-second intervals since the last
measurement reset that contained excess zeros.
All 1's - A tally of the number of one-second intervals since the last
measurement reset that contained the AIS pattern.
Error Free Seconds - A tally of the number of one-second intervals that were
error free since the last measurement reset.
Errored Seconds - A tally of the number of one-second intervals since the last
measurement reset that contained errors such as BPVs and frame errors.
Severe Errored Seconds - A tally of the number of one-second intervals since
the last measurement reset that were severely errored.
Available Seconds - A tally of the number of one-second intervals since the last
measurement reset that were available for service.
Unavailable Seconds - A tally of the number of one-second intervals since the
Vista de página 318