Agilent Technologies Agilent B2201A Manual do Utilizador Página 147

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 234
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 146
Agilent B2200 Users Guide, Edition 2 5-21
SCPI Command Reference
:DIAGnostic:TEST:CARD:CLEar
:DIAGnostic:TEST:CARD:CLEar
This command clears the relay test result (pass/fail result) of the specified card.
Syntax :DIAGnostic:TEST:CARD:CLEar card_number
Example OUTPUT @Agb2200;":DIAG:TEST:CARD:CLE 1"
:DIAGnostic:TEST:CARD[:EXECute]?
This command executes the relay test for specified card, then returns the pass/fail
result. Before starting the relay test, open the input/output terminals (end of cable is
OK, Kelvin cable must be removed). For more information, see “RELAY_TEST”
on page 3-35.
NOTE This command changes all settings of card to same state as after executing
:SYST:CPON command.
Syntax :DIAGnostic:TEST:CARD[:EXECute]? card_number
Query response test_result <newline><^END>
1: failure card exists
0: pass
Example OUTPUT @Agb2200;":DIAG:TEST:CARD? ALL"
Parameter Explanation
card_number card number: 1, 2, 3, 4, or ALL
Parameter Explanation
card_number card number: 1, 2, 3, 4, or ALL
Vista de página 146
1 2 ... 142 143 144 145 146 147 148 149 150 151 152 ... 233 234

Comentários a estes Manuais

Sem comentários