Agilent Technologies Agilent B2201A Manual do Utilizador Página 114

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 234
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 113
4-16 Agilent B2200 Users Guide, Edition 2
Programming
Programming Examples
Saving Input/Output Labels
You can define labels to use for the input/output ports when you control the switch
in the GPIB local mode. The labels are cleared by *RST. So it is recommended to
save the labels into the internal memory and define a comment for the memory data.
You can see the comment when you load/save the memory data in the GPIB local
mode.
Setup: Channel configuration mode: Auto
Label for input 1: SMU1
Label for input 2: SMU2
Label for input 3: SMU3
Label for input 4: UNUSED
Label for input 5: SMU4-F
Label for input 6: SMU4-S
Label for input 7: SMU5-F
Label for input 8: SMU5-S
Label for input 9: OPEN
Label for input 10: BIAS
Label for input 11: OPEN
Label for input 12: GROUND
Label for input 13: CMU-H
Label for input 14: CMU-L
Label for output 1: BULK
Label for output 2: SOURCE
Label for output 3: GATE
Label for output 4: DRAIN
Label for output 5: GROUND
Comment for memory 1: Port label info
Vista de página 113
1 2 ... 109 110 111 112 113 114 115 116 117 118 119 ... 233 234

Comentários a estes Manuais

Sem comentários