Agilent Technologies Agilent B2201A Manual do Utilizador Página 139

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 234
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 138
Agilent B2200 Users Guide, Edition 2 5-13
SCPI Command Reference
Command Summary
Subsystem Commands
Command Summary
ROUT subsystem :ROUTe subsystem has commands for controlling the signal routing.
Open a channel”: opens relays to disconnect the channel (that is, disconnects input
port from output port).
Close a channel”: closes relays to connect the channel (that is, connects input port
to output port).
Command Description
[:ROUT]:FUNC channel_config
[:ROUT]:FUNC?
Sets the channel configuration mode.
channel_config:
ACON: Auto Config Mode
NCON: Normal Config Mode
Query returns the present channel configuration:
ACON or NCON.
[:ROUT]:CONN:RULE card_number,rule
[:ROUT]:CONN:RULE? card_number
Sets the connection rule for the specified card.
card_number: 0 or ALL for Auto Config,
1, 2, 3, 4, or ALL for Normal Config
rule: FREE (free) or SROUte (single)
Query returns the connection rule of the specified
card: FREE or SROU.
card_number: Card to check. 0 for Auto Config,
1, 2, 3, or 4 for Normal Config.
[:ROUT]:CONN:SEQ card_number,sequence
[:ROUT]:CONN:SEQ? card_number
Specifies the connection sequence mode for the
specified card.
card_number: 0 or ALL for Auto Config,
1, 2, 3, 4, or
ALL for Normal Config
sequence:
NSEQ: No-Sequence mode
BBM: Break-Before-Make mode
MBBR: Make-Before-Break mode
Query returns the connections sequence mode of
the specified card: NSEQ, BBM, or MBBR.
card_number: Card to check. 0 for Auto Config,
1, 2, 3, or 4 for Normal Config.
Vista de página 138
1 2 ... 134 135 136 137 138 139 140 141 142 143 144 ... 233 234