Chapter 3 Programming Your Counter for Remote Operation
Status Reporting
3-30 Programming Guide
3
Standard Event Status Register Group
Figure 3-12. Standard Event Status Reporting
Standard Event Status Register
The Standard Event Status Register contains bits that monitor specific
IEEE 488.2-defined events as shown in Figure 3-12.
Use *ESR? to read this register.
Use *ESR? or *CLS to clear this register.
Table 3-5 lists the Standard Event Status Register bits and briefly
describes each bit.
0
1
2
3
4
5
6
7
&
&
&
&
&
&
&
0
1
2
3
4
5
7
Standard Event
Status Register
To ESB bit (bit 5) of the
Status Byte Register
Logical OR
6
Operation Complete
Not Used
Query Error
Device-Dependent Error
Execution Error
Command Error
Power On
Not Used
&
Standard Event
Enable Register