Chapter 4 Command Reference
[:SENSe] Subsystem
Programming Guide 4-25
4
COMMAND [:SENSe]:FREQuency:OFFSet . . . <numeric_value>[<frequency unit>]
Sets a reference frequency for all other absolute frequency settings in
the instrument.
RANGE The acceptable range for the <numeric_value> parameter is 0 to 50 GHz.
UNITS The offset frequency can be specified in Hz, KHz, or MHz only.
COMMENTS This command does not affect the hardware settings of the instrument.
It affects only the entered and displayed frequencies.
The Counter accepts only six digits of resolution for an offset entry
through this command. For example, the command
FREQ:OFFSet 12345.678912MHz results in an offset of approximately
12.3456 GHz.
The coupling equation for this command is:
Entered | Displayed frequency = (Hardware frequency ) + offset.
RELATED
FRONT-PANEL
KEYS
Freq Offset
COMMAND [:SENSe]:FREQuency:OFFSet:STATe . . . <Boolean>
When FREQuency:OFFSet is ON, the frequency measurement results
are modified by the setting of FREQ:OFFset before being displayed or
reported (in response to a query).
RELATED
FRONT-PANEL
KEYS
Freq Offset