Agilent 75000 SERIES C
Agilent E1411A/E1411B
5
1
/
2
-Digit Multimeters
Service Manual
Serial Numbers
The Information in this Service Manual applies to E1411A 5 1/2-Digit
Multimeters with serial number prefix 3002A and to E1411B 5 1/2-Digit
Multimeters with serial number prefix 3050A.
*E1411-90011*
E1411-90011
Copyrigh Agilent Technologies, Inc. 1996-2005
Manual Part Number: E1411-90011 Printed: November 2005 Edition 2
Printed in U.S.A. E1105
Vista de página 0
1 2 3 4 5 6 ... 79 80

Resumo do Conteúdo

Página 1 - *E1411-90011*

Agilent 75000 SERIES CAgilent E1411A/E1411B 5 1/2-Digit MultimetersService ManualSerial Numbers The Information in this Service Manual applies

Página 3 - Table of Contents

Chapter 1General InformationIntroduction This manual contains information required to test, adjust, troubleshoot, andrepair the Agilent E1411A and E14

Página 4

Warnings andCautions This section contains WARNINGS which must be followed for yourprotection and CAUTIONS which must be followed to avoid damage to

Página 5

WARNINGUSE PROPER FUSES. For continued protection against firehazard, replace the line fuse(s) only with fuses of the samecurrent rating and type (su

Página 6 - Safety Symbols

Multimeter Serial NumbersMultimeters covered by this manual are identified by a serial number prefixlisted on the title page. Agilent Technologies use

Página 7 - DECLARATION OF CONFORMITY

Chapter 2InstallationIntroduction This chapter provides information to install the E1411 multimeter, includinginitial inspection, preparation for use,

Página 8

Shipping the MultimeterIf you need to return an E1411 multimeter to Agilent, first remove anyadapters or connectors before packaging the instrument f

Página 9

Chapter 3 Operating InstructionsIntroduction This chapter lists operating information for the Agilent E1411 multimeter,including:• Multimeter operatio

Página 10

NOTETest failures can be caused by improper cabling, improper selection of theinterface select code, primary, and/or secondary address setting. Verif

Página 11 - General Information

Chapter 4Verification TestsIntroduction The three levels of test procedures described in this chapter are used to verify that the E1411 multimeter:• i

Página 13 - Description

The value in the "Measurement Uncertainty" column of Table 4-1 isderived from the specifications for the source used for the test, andrepre

Página 14 - Ω to 1 MΩ

NOTETest failures can be caused by improper selection of the interface selectcode, primary address setting, and/or secondary address setting. Verifyp

Página 15 - Chapter 2

Test 4-1: DC Voltage Test (Zero Volt Input)This test verifies DC Voltage accuracy on all five ranges using a zero volt input.Equipment Setup 1. Conn

Página 16 - Shipping

Test Procedure 1. Set the E1411 range to 0.113 V (0.125 V with 10% over- range) andmeasure the input voltage with MEAS:VOLT:DC? 0.12. Observe the in

Página 17 - Operating Instructions

Test 4-2: DC Voltage Test (DCV Input)This test verifies DC Voltage accuracy on all five ranges with DC voltageinputs.Equipment Setup 1. Connect the e

Página 18 - Self-Test

NOTE*RST sets Auto Zero to ON and Power Line Cycles to 1. Test Procedure 1. Set the DC Standard Output to 0.1 DCV.2. Set the E1411 range to 0.113 V

Página 19 - Verification Tests

90 FOR I=1 TO 5100 PRINT "Set DC Standard to ";Volts(I);"VDC"110 PRINT "Press Continue when ready"1

Página 20 - Verification

2. Set the E1411 multimeter as follows:• Reset Multimeter ...*RST • Auto Zero ...

Página 21 - Example: Self-Test

E1411Range10%overrangeAC Standard OutputVoltage (Vac) Frequency (Hz)79.5 mV79.5 mV79.5 mV300.0 V87.5 mV87.5 mV87.5 mV None0.07 V0.07 V0.07 V300.00

Página 22

230 ENDTest 4-4: Resistance Test (4-Wire Ohms)This test verifies the 4-wire resistance accuracy of the 2kΩ, 100kΩ, and1MΩ ranges.NOTEThe DC Voltage p

Página 23 - DCV Test

Table of ContentsChapter 1 - General InformationIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 24

2. Set the E1411 multimeter as follows:• Reset Multimeter ...*RST• Auto Zero ...

Página 25 - Test (DCV Input)

Example: 4-Wire Ohms TestThis example performs a 4-wire ohms resistance test for a power linereference frequency of 60 Hz. Change line 80 to OUTPUT709

Página 26 - Test 4-3: AC Voltage Test

Performance Test RecordTable 4-1, Performance Test Record for the E1411 Multimeter, is a formyou can copy and use to record performance verification t

Página 27

Table 4-1. Performance Test Record for the E1411 Multimeter (Page 1 of 3)Test Facility:Name _________________________________________Address _______

Página 28 - AC Voltage

Table 4-1. Performance Test Record for the E1411 Multimeter (Page 2 of 3)Model _________________________________ Report No. ____________________

Página 29 - 230 END

Table 4-1. Performance Test Record for the E1411 Multimeter (Page 3 of 3)Model ________________________________ Report No. ________________________

Página 30 - Test Procedure 1

34 Verification Tests Agilent E1411A/E1411B Service Manual

Página 31 - Ohms Test

Chapter 5AdjustmentsIntroduction This chapter contains procedures to adjust the Agilent E1411 multimeter forpeak performance. For best performance, t

Página 32 - Test Record

DC VoltageAdjustments This procedure adjusts E1411 multimeter DC voltage measurementaccuracy.Equipment Setup1. Connect the equipment as shown in Figu

Página 33

NOTE*RST sets Auto Zero to ON and Power Line Cycles to 1.AdjustmentProcedure1. Set the DC Standard output to 7.7 V2. Set the E1411 range to 8 V, and

Página 34 - Test Equipment Used:

Chapter 6 - Replaceable PartsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 35

When connected to the E1411 multimeter, some DC Standards may exhibitvoltage variations at the start of a measurement. TRIG:DEL .05 (line 190)adds a 0

Página 36

AC VoltageAdjustmentsThis procedure adjusts the E1411 AC voltage measurement accuracy.NOTEThe DC Voltage adjustment MUST be performed before the AC V

Página 37 - Adjustments

NOTE*RST sets Auto Zero to ON and Power Line Cycles to 1.AdjustmentProcedure1. Set the AC Standard output to 5.6 Vac at 1 kHz2. Set the E1411 range

Página 38 - DC Voltage

1 ! AC Voltage Adjustments10 OUTPUT 70903;"*RST" !Set autozero on and PLC 120 OUTPUT 70903;"CAL:LFR 60" !Set 60 Hz line re

Página 39 - Example: DC

ResistanceAdjustmentsThis procedure adjusts the E1411 multimeter 4-wire resistance measurementaccuracy.NOTEThe DC Voltage adjustment and the AC Vo

Página 40

NOTE*RST sets Auto Zero to ON and Power Line Cycles to 1.AdjustmentProcedure1. Set the Resistance Standard to 1 kΩ.2. Set the E1411 range to 2 kΩ, a

Página 41 - AC Voltage

Example: 4-WireResistanceAdjustmentsThis example performs a 4-wire ohms resistance adjustment for a powerline reference frequency of 60 Hz. Change lin

Página 42

CalibrationErrorsTable 5-1 summarizes calibration error numbers, titles and descriptions forthe E1411 multimeter. Note that an error is returned if t

Página 43

46 Adjustments Agilent E1411A/E1411B Service Manual

Página 44 - Resistance

Chapter 6Replaceable PartsIntroduction This chapter contains information to order replaceable parts for the AgilentE1411A and E1411B multimeters. Exch

Página 45

CertificationAgilent Technologies, Inc. certifies that this product met its published specifications at the time of shipment from the factory. Agilent

Página 46

Table 6-1. E1411A Replaceable PartsReference*DesignatorPartNumberQty Description Mfr**CodeMfr Part NumberE1411-66201E1411-6920111EXCHANGE ASSEMBLIESE1

Página 47 - Calibration

Figure 6-1. Agilent E1411A Replaceable PartsAgilent E1411A/E1411B Service Manual Replaceable Parts 49

Página 48

Table 6-2. Agilent E1411B Replaceable PartsReference*DesignatorPartNumberQty Description Mfr**CodeMfr Part NumberE1411-66511***E1411-69511***11EXCHANG

Página 49 - Replaceable Parts

Figure 6-2. Agilent E1411B Replaceable PartsAgilent E1411A/E1411B Service Manual Replaceable Parts 51

Página 50

Table 6-3. E1411 Reference DesignatorsE1411 Reference DesignatorsA ... assemblyCBL ...

Página 51

Chapter 7Manual ChangesIntroduction This chapter contains information to adapt this manual to instruments forwhich the content does not directly apply

Página 52

54 Manual Changes Agilent E1411A/E1411B Service Manual

Página 53

Chapter 8ServiceIntroduction This chapter contains information to service the E1411 multimeter includingtroubleshooting guidelines and repair/maintena

Página 54

TroubleshootingTechniques There are two main steps to troubleshoot an E1411 multimeter problem:(1) identify the problem, and (2) test assemblies to

Página 55 - Manual Changes

Performance Out of Specification If the multimeter performance is out of specification limits, use theadjustment procedures in Chapter 5 - Adjustments

Página 56

Frame or chassis ground terminal—typi-cally connects to the equipment’s metalframe.Alternating current (AC).Direct current (DC). Indicates hazardous v

Página 57 - Chapter 8

Checking Switches/JumpersVerify the logical address setting is set correctly (factory set at 24). Verifythe interrupt priority jumpers are set correct

Página 58 - Techniques

• Do not use pliers to remove a MOS or CMOS device from ahigh-grip socket. Instead, use a small screwdriver to pry the deviceup from one end. Slowly l

Página 59 - Assemblies

NOTE When reinstalling the top shield, verify that the slots are aligned with thetabs, then set the top shield into place.Removing Front PanelUse the

Página 60 - Guidelines

Removing Binding PostsUse the following steps to remove the E1411 front panel binding posts(MP3, MP4, MP5, and MP6) (see Figure 8-3).1. Unsolder wir

Página 61 - Top Shield

• Avoid unnecessary component unsoldering and soldering. Excessivereplacement can result in damage to the circuit board and/oradjacent components.• Do

Página 62 - Front Panel

Appendix ACalculating Multimeter AccuracyIntroduction This appendix shows how multimeter accuracy, measurement uncertainty,and test accuracy ratio (T

Página 63 - Printed Circuit

Test Accuracy Ratio (TAR) DefinitionTest Accuracy Ratio (TAR) is the ratio of multimeter accuracy tomeasurement uncertainty. For the E1411 multimeter

Página 64 - Safety Checks

AC VoltageAccuracy EquationsFrom Appendix A- Specifications of the E1326A /E1411A or E1326B/E1411B User’s Manual, AC voltage 90-day accuracy = ±(% of

Página 65 - Appendix A

Measurement UncertaintyCalculationsMeasurement uncertainties for the Datron 4708 source are calculated usingthe 90-day accuracy specifications in the

Página 66 - Calculations

Calculate ACVMeasurement Uncertainty From Section 6 - Specifications of the Datron 4708 User’s Handbook , ACVoltage (Option 20) Accuracy (90 days sinc

Página 67

SADECLARATION OF CONFORMITYAccording to ISO/IEC Guide 22 and CEN/CENELEC EN 45014Revision: A.03 Issue Date: 09/05/00Manufacturer’s Name:Agilent Techno

Página 68 - Uncertainty

Test AccuracyRatio (TAR)Calculations For the E1411 multimeter, Test Accuracy Ratio (TAR) = [High Limit -Input Value]/Measurement Uncertainty where

Página 69

Appendix BVerification Tests - C ProgramsFunctionalVerificationTestThis program is designed to do the Functional Verification Test found inChapter 4 -

Página 70 - Ratio (TAR)

PerformanceVerificationTestsThese programs are designed to do the Performance Verification Testsfound in Chapter 4 - Verification Tests.Example: Zero

Página 71 - Appendix B

Example: DCVoltage TestThis test performs a DC Voltage test for positive input DC volts and apower line reference frequency of 60 Hz. /* DC Voltage T

Página 72

Example: ACVoltage TestThis example performs an AC voltage test for a power line referencefrequency of 60 Hz./* AC Voltage Test E1411A/B *

Página 73 - Voltage Test

Example:Resistance TestThis example performs a 4-wire ohms resistance test. The program alsocalculates the Upper and Lower Limit values for the ACTUA

Página 74 - Example: AC

Adjustments These programs are designed to do the adjustments found in Chapter 5 -Adjustments.DC VoltageAdjustmentsThis example performs DC Voltage ad

Página 75 - Resistance Test

AC VoltageAdjustmentsThis example performs an AC Voltage adjustment for a power linereference frequency of 60 Hz and an input of 5.6 Vac at 1 kHz. If

Página 76

ResistanceAdjustmentsThis example performs a 4-wire ohms resistance adjustment for a powerline reference frequency of 60 Hz. If a calibration error o

Página 79

Copyright © Agilent Technologies, Inc. 1996-2005