Agilent Technologies 4294A Manual de Instruções Página 83

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 469
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 82
Chapter 4 75
Preparation of Measurement Accessories
Adapter Setting
4. Preparation of
Measurement Accessories
KIT” is displayed in the instrument state area in the LCD's upper-left side.
Step 1. Make sure that the 42942A is correctly connected to the four-terminal pair port of the
Agilent 4294A.
Figure 4-4 Connecting the 42942A
NOTE The 42942A must be directly connected to the four-terminal pair port of the Agilent
4294A. If another accessory such as a cable or adapter is inserted between the 42942A and
the four-terminal pair port of the Agilent 4294A, the correct adapter setup data cannot be
obtained and correct device measurement cannot be performed after the adapter setup.
Step 2. More than 30 minutes warm-up time is required after turning on the Agilent 4294A.
Step 3. Make the correct adapter selection for the 42942A in accordance with “Adapter Selection”
on page 68.
Step 4. After performing Step 4 of “Adapter Selection” on page 68, press the
SETUP key to
display the Adapter Setup Menu.
Step 5. Connect OPEN (0 S) to the 7-mm port of the 42942A (the OPEN standard is furnished with
the 42942A Terminal Adapter).
Vista de página 82
1 2 ... 78 79 80 81 82 83 84 85 86 87 88 ... 468 469

Comentários a estes Manuais

Sem comentários