Agilent Technologies 4294A Manual de Instruções Página 273

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 469
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 272
Chapter 8 265
Analysis and Processing of Result
Set a limit to the trace and make pass/fail evaluation
8. Analysis and Processing
of Result
Figure 8-29 Limit line movement by offset (+ 1 kHz)
Step 6. Press the
AMPLITUDE OFFSET key to display the current measurement parameter offset
for the limit line in the set parameter value field in the upper left area of the screen.
Step 7. Use the keys or rotary knob of the ENTRY block in one of the following ways to specify
the measurement parameter offset value.
Enter the desired value with the numeric keys (
[0] to [9], [-]and [.]) and then press one
of the unit keys (
[G/n], [M/m], [k/m], or [´1]).
Turn the rotary knob (
m
mm
m) until the desired value is set.
Press the step keys ([][¯]) to set the desired value.
Vista de página 272
1 2 ... 268 269 270 271 272 273 274 275 276 277 278 ... 468 469

Comentários a estes Manuais

Sem comentários