330 Chapter 10
Specifications and Supplemental Performance Characteristics
Basic Characteristics
Adapter Setup
Calibration
Measurement Accuracy
Accuracy Specifications
Averaging factor 1 to 256 (integer)
Adapter Selection
NONE No adapter (the 16047E or other direct connection
type test fixture) is connected
4TP 1M Four-terminal pair 1 m (16048G)
4TP 2M Four-terminal pair 2 m (16048H)
7-mm 42942A Terminal Adapter (42942A)
PROBE 42941A Impedance Probe (42941A)
Calibration
User calibration Calibration performed with user-defined calibration
kit (OPEN, SHORT, LOAD)
Port extension Compensation performed when the measurement
terminal is expanded from the 7-mm connector of the
42942A Terminal Adapter or the 3.5-mm connector
of the 42941A Impedance Probe. Enter electrical
length or delay time for the extension.
Fixture compensation Compensation performed at the device contacts of the
test fixture using OPEN, SHORT, LOAD.
Calibration points Fixed points or user points determined by sweep
setups
Temperature
Four-terminal pair port of
the 4294A’s front panel
23 ± 5°C
Beyond 23 ± 5°C, the measurement accuracy
decreases to half the value described.
7-mm port of the
42942A Terminal
Adapter
Within ± 5°C from the adapter setup temperature.
Measurement accuracy applies when the adapter
setup is performed at 23 ± 5°C. When the adapter
setup is performed beyond 23 ± 5°C, the
measurement accuracy decreases to half the value
described.