226 Chapter 8
Analysis and Processing of Result
Search the point of target measurement
Figure 8-8 Result of target value search (Trace B)
NOTE If multiple target values are present on the trace, the main marker moves to the target value
of the measuring points that were closest to the main marker immediately before the target
value search was executed.
If the target value is not found, the message “TARGET VALUE NOT FOUND” is
displayed in the upper left area of the screen.
Step 5. If multiple target measurements are present on the trace, operate the following key to move
from the first searched point to other points.
Search type Key operation
Search in the left direction (search from the
current marker position in the direction of
decreasing sweep parameter values, and move
the main marker to the first detected target
value)
SEARCH LEFT
Search in the right direction (search from the
current marker position in the direction of
increasing sweep parameter values, and move
the main marker to the first detected target
value)
SEARCH RIGHT