11 Specifications and Characteristics
260 N1911A/1912A P-Series Power Meters User’s Guide
Measurement Speed
Pulse Characteristic
These pulse characteristics are only applicable when a P- Series wideband
sensor is used.
5
Applies to frequencies >500 MHz (See Figure 11-158).
6
Applies only when Off is selected for video bandwidth.
7
The Minimum Pulse Width is the recommended minimum pulse width viewable on the power
meter, where power measurements are meaningful and accurate, but not warranted.
8
Based on 10 samples per period.
Maximum Sample Rate and Capture Length
Measurement speed using remote
interface
1500 readings per second
Rise Time and Fall Time
13 ns
5,6
Minimum Pulse Width
50 ns
7
Overshoot
5 %
Maximum Pulse Repetition Rate
10 MHz
8
Maximum Sampling Rate
100 Msamples/second,
continuous sampling
Maximum Capture Length
1 second