Using E9320 E-Series Power Sensors 4
N1911A/1912A P-Series Power Meters User’s Guide 147
Improving the Accuracy with Additional Settings
You can now return to the setup menus and adjust the default setting to
obtain greater measurement accuracy.
Channel Setup
Press . The Channel Setup screen is displayed. If using an
N1912A, select the channel you want to configure.
Figure 4-111E-Series E9320 power sensor dafault channel setup
Use the , , , and keys to highlight the setting field and press
to configure the setting. For further detail on these settings, refer
to the appropriate sections.
Meas Avg
: “Setting Measurement Averaging” on page 49
Video Avg
: “Setting the Video Averaging” on page 52
Step Detec
t: “Step Detection” on page 51
Video B/W: “Setting the Video Bandwidth” on page 54
NOTE
For range settings, only UPPER and LOWER ranges are available in Normal and Triggered
mode.