Agilent Technologies 1912A Guia do Utilizador Página 232

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 297
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 231
8 Using N8480 Series Power Sensors
206 N1911A/1912A P-Series Power Meters User’s Guide
Figure 8-140Calibration factor displayed
Example
To make a measurement on channel A with a power sensor which has a
reference calibration factor of 99.8% and a calibration factor of 97.6% at
the measurement frequency.
• Disconnect the power sensor from any signal source.
Press ,
and the channel softkey.
• Use the numeric keypad to enter 99.8 in the Ref Cal Factor pop- up
window.
Press
to complete the entry.
Press and the channel softkey. The Zeroing pop- up is
displayed.
Connect the power sensor to the POWER REF output.
Press and the channel softkey to start the calibration
routine. The Calibrating pop- up is then displayed.
Press , . The value is displayed on the Cal Fac field.
Calibration Factor
REF CFs
REF CF
%
Zero
Cal
Offset
Vista de página 231
1 2 ... 227 228 229 230 231 232 233 234 235 236 237 ... 296 297

Comentários a estes Manuais

Sem comentários