Using 8480 Series Power Sensors 7
N1911A/1912A P-Series Power Meters User’s Guide 175
Power Meter Configuration
The P- Series power meters automatically recognize an 8480 Series power
sensor when it is connected. The averaging settings shown in Figure 7- 123
are automatically configured.
Figure 7-1238480 Series auto-averaging settings
10 dB
10 dB
Maximum Sensor Power
Minimum Sensor Power
Power Sensor
Dynamic Range
10 dB
10 dB
1234
1118
Resolution Setting
Number of Averages
11116
11232
1 1 16 256
1 8 128 128
NOTE
These values are valid only for the power meter channel connected with Agilent 8480
Series power sensors. Averaging settings can also be manually configured.