2 General Power Meter Functions
108 N1911A/1912A P-Series Power Meters User’s Guide
Press ,
and toggle the to On.
The Zeroing pop- up is displayed when a zero occurs.
Manual Zeroing
To manually zero the power meter and sensor:
Press and the channel softkey.
The Zeroing pop- up is displayed.
Calibration
Calibration sets the gain of each power meter channel and P- Series sensor
combination. This is achieved without the need to connect it to the
1.0 mW Power Reference. The power meter can be set to automatically
calibrate on- the- fly while you are using it. Alternatively, you can perform
the calibration manually. The reference calibration factor is automatically
set for all P- Series power sensors.
During calibration, the wait symbol is displayed.
Figure 2-88 Calibration wait pop-up
1 of 2
Auto Zero A
NOTE
When you are performing measurements, this may cause delays in obtaining results.
During this time, it is recommended to disable the automatic zeroing.
Zero
NOTE
On dual channel meters you can zero each channel independently or both channels
sequentially by pressing the softkey - choosing , or
Zero
Zero A
Zero B
Zero Both