Agilent Technologies HP 8566B Especificações Página 52

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 405
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 51
8. Amplitude Fidelity Test
8.
Amplitude
Fidelity
Test
(For instruments with Option 857, refer to Chapter 5.)
Related
Adjustment
Log Amplifier Adjustments
Specification
Log:
Incremental
fO.l
dB/dB
over 0 to 80
dB
display
Cumulative
3 MHz to 30 Hz Resolution Bandwidth
s&l.0
dB
over 0 to 80
dB
display (20 to 30°C)
sf1.5
dB
over 0 to 90
dB
display Linear:
4~3%
of Reference Level for top 9 l/2 divisions of display
Description
Amplitude fidelity in log and linear modes is tested by decreasing the
signal level to the spectrum analyzer in 10
dB
steps with a calibrated
signal source and measuring the displayed amplitude change with the
analyzer’s MARKER A function.
FREOUENCY
SYNTHESIZER
SPECTRUM ANALYZER
TER
Figure 2-11. Amplitude Fidelity Test Setup
Performance Tests 2-27
Vista de página 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 404 405

Comentários a estes Manuais