Agilent Technologies HP 8566B Especificações Página 265

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 405
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 264
13. Sweep, DAC, and Main Coil Driver Adjustments
73. Use signal generator TUNING control to vary signal generator
output frequency
flO0
MHz (from frequency in step 70) in 1 MHz
steps. Readjust
A19R43
25
GHz
SPAN slightly as necessary so that
amplitude of displayed signal peak varies less than 3
dB
over
entire
flO0
MHz range bracketing band overlap point.
74. Key in
(SHIFT),
[PRESEL
PEAK)=.
75. Verify that jumper between
A12TP2
and
A12TP3
(Lock Indicator
Disable) has been removed. Replace RF Digital Section cover over
Al2 through Al6 assemblies.
Adjustments 3-125
Vista de página 264
1 2 ... 260 261 262 263 264 265 266 267 268 269 270 ... 404 405

Comentários a estes Manuais

Sem comentários