2 1. Frequency Response Adjustments
126.
127.
128.
129.
the high end of the band drop in amplitude by approximately
0.75
dB.
Readjust 22-turn (IF Offset) potentiometer
A6A12R83
D as
necessary to maximize the overall level of the displayed signal
responses from 12.5
GHz
to 18.6
GHz
on the spectrum analyzer
display. If the displayed signal responses peak at widely different
settings of
A6A12R83
D, perform the following steps:
a. Adjust 22-turn (YTX Linearity) potentiometers
A6AlOR41
LB2,
A6AlOR42
LB3,
A6AlOR70
LB4,
A6A610R34
LR2,
A6AlOR37
LR3, and
A6AlOR76
LR4 fully counterclockwise
if they have not already been so adjusted; do not readjust
A6A12R40
LB1 or
A6AlOR31
LRl.
b. Readjust 22-turn (IF Offset) potentiometer
A6A12R83
D as
necessary to maximize the overall level of the displayed signal
responses at the low end of the band.
c. On the spectrum analyzer, key in MARKER
(j-1
and
position the marker at the point in the band where the signal
responses begin to fall off from their maximum value.
d. Adjust 22-turn (YTX Linearity) potentiometer
A6AlOR41
LB2
clockwise until the displayed signal responses at the high end
of the band begin to drop. Continue to adjust
A6AlOR41
LB2
until the
rolloff
point aligns with the position of the marker.
See Figure 3-90 for the location of
A6AlOR41
LB2,
A6AlOR42
LB3,
A6AlOR34
LR2 and
A6AlOR37
LR3.
e. Readjust 22-turn (YTX Linearity) potentiometer
A6AlOR34
LR2 clockwise to maximize the displayed signal responses for
approximately 3 divisions to the right of the marker.
f. On the spectrum analyzer, key in MARKER
[NORMAL)
and
position the marker at the point in the band where the signal
responses begin to fall off from their maximum value.
g. Readjust 22-turn (YTX Linearity) potentiometer
A6AlOR42
LB3 clockwise until the displayed signal responses at the high
end of the band begin to drop. Continue to adjust
A6AlOR42
LB3 until the
rolloff
point aligns with the position of the
marker.
h. Readjust 22-turn (YTX Linearity) potentiometer
A6AlOR37
LR3 clockwise to maximize the displayed signal responses at
the high end of the band.
i. Readjust
A6A12R83
D,
A6AlOR34
LR2, and
A6AlOR37
LR3
as necessary to maximize the displayed signal responses from
12.5
GHz
to 18.6
GHz
on the spectrum analyzer display.
Gradually readjust (Band D Lower Segment) A6AllR57
Dl
and (Band D Upper Segment) A6AllR75 D2 as necessary for
maximum flatness of the displayed signal responses.
On the spectrum analyzer, key in SWEEP
(SINGLE),
[SWEEP TIME)
15Os,
TRACE A [BLANK), TRACE B
[MAX]),
I-1
6, TRACE B
[CLEAR-WRITE],
lsAVEJ
5,
m).
Press SWEEP
CRINGLE]
on the spectrum analyzer and wait for the
sweep to complete (150 seconds) and the SWEEP LED to turn
off. As the spectrum analyzer tunes from 12.5
GHz
to 18.6
GHz,
the spectrum analyzer frequency response should be displayed as
TRACE B.