Agilent Technologies B1500A Guia do Utilizador Página 68

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 204
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 67
2- 26 Keysight B1500A User’s Guide, Edition 14
Introduction
Measurement Modules
MCSMU - Medium Current SMU
This section describes typical specification of the medium current source/monitor
unit (MCSMU). Maximum four modules can be installed in one mainframe.
Maximum current: ± 1 A (pulse), ± 0.1 A (DC)
Maximum voltage: ± 30 V
Maximum output power: 30 W (pulse), 3 W (DC)
Minimum range: 0.2 V, 10 A
Minimum pulse width: 10 s
Output/measurement value and resolution: see Table 2-9 and Table 2-10.
Figure 2-5 MCSMU Output and Measurement Ranges
30
30
1
0.1
0.1
1
Voltage (V)
Current (A)
-
-
-
Pulse mode only
Pulse mode only
Vista de página 67
1 2 ... 63 64 65 66 67 68 69 70 71 72 73 ... 203 204

Comentários a estes Manuais

Sem comentários