3- 42 Keysight B1500A User’s Guide, Edition 14
Installation
Mounting Connectors
To Connect MFCMU Output
The MFCMU output connection example is shown in Figure 3-21. This connection
is for the MFCMU without the SCUU. Then Keysight N1300A-001/002 CMU cable
should be used for the connection between the MFCMU and the connectors you
mount. It is ok to leave open the ground wire extended from the CMU cable.
Figure 3-21 MFCMU Output Connection
The Hcur/Hpot/Lcur/Lpot lines should be connected to the DUT as shown above.
Use a 50 coaxial cable (Keysight part number: 8120-0367) for the
Hcur/Hpot/Lcur/Lpot lines connection. Use a wire for the guard connection.
Hcur
Guard
Hpot
Lpot
Guard
Lcur
Insulator
Plate
BNC connector
Coaxial cable
Wire
to DUT 1
high terminal
to DUT 1
low terminal