Keysight B1500A User’s Guide, Ed ition 14 3- 45
Installation
Connecting Measurement Devices
Connections for High Current Measurements
When you force or measure a large current, you may want to use a Kelvin
connection (4-wire, remote sensing) to eliminate the residual resistance effects of
test leads and contacts. For example, you can use the following connections as
Kelvin connections on the test fixture.
Examples: Kelvin Connection
To cancel the effects of the residual resistance, test leads must be connected as close
as possible to the DUT.
SMU(±200V= 1A Max)
SenseForce
Guard Guard
1
SenseForce
Guard Guard
2
SenseForce
Guard Guard
3 (Selected)
Gate
Source
Drain
Substrate