CRT Monitor
Assembly
Adjustments
Intensity,
Sub-bright,
and Contrast
Adjustment
1. Refer to figure for adjustment locations.
2. Place TIMING WAVEFORMS menu on the screen of the HP 1650B/51B, by
pressing front-panel DISPLAY key.
Note This menu is used because it has characters throughout the screen which are
watched during the procedures. Any other menu may be used; however, the
adjustments may not be as accurate if characters and/or lines are not displayed
throughout the screen.
3. Set rear-panel INTENSITY to the minimum setting.
4. Adjust sweep board SUB BRIGHT control to the lowest setting of brightness
where menu is visible on the CRT screen.
Caution High voltages exist on the sweep board. Avoid contact with the sweep board when
making CRT adjustments.
Figure 4-2. CRT Adjustment Locations
Adjustments HP 1650B/1651B
4-4 Service Manual