Agilent Technologies 54502A Manual de Serviço Página 59

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 111
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 58
Glitch Test Description:
This performance test verifies the glitch detection specification of the
HP 1650B/51B.
Specification:
Minimum detectable glitch: 5 ns wide at the threshold.
Equipment:
Pulse Generator ............................... HP 8161A/020
Oscilloscope...................................... HP 54502A
50 ohm Feedthrough.(1)............................ HP 10100C
Test Connector (1) ...........................see figure 3-1
Procedure:
1. Connect the test equipment as in figure 3-29. The clock inputs are not used
for the glitch test since glitch detection is part of timing analysis. Use the
oscilloscope to make sure pulses are 5 ns wide at the threshold (1.6 V).
Note In this setup, only eight channels are tested at a time to minimize loading. Ground
lead must be grounded to ensure accurate test results.
Figure 3-29. Setup for Glitch Test
Performance Tests HP 1650B/1651B
3-24 Service Manual
Vista de página 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 ... 110 111

Comentários a estes Manuais

Sem comentários